NABOK, A. V., TSARGORODSKAYA, A., HOLLOWAY, A., STARODUB, N. F. and GOJSTER, O. (2007). Registration of T-2 mycotoxin with total internal reflection ellipsometry and QCM impedance methods. Biochemical and Biomolecular Sensing, 22 (6), 885-890. [Article]
Abstract
A sensitive optical method of total internal reflection ellipsometry (TIRE) in conjunction with immune assay approach was exploited for the registration of T-2 mycotoxin in a wide range of concentrations from 100 mu g/mI down to 0.15 ng/ml. Association constants of 1.4 x 10(6) and 1.9 x 10(7) mol(-1) s for poly- and monoclonal T-2 antibodies, respectively, were evaluated from TIRE kinetic measurements. According to TIRE data fitting, binding of T-2 molecules to antibodies (at saturation) has resulted in the increase in adsorbed layer thickness of 4-5 nm. The QCM impedance measurements data showed anomalously large mass increase and film softening, most likely, due to the binding of large T-2 aggregates to antibodies. (c) 2006 Elsevier B.V. All rights reserved.
More Information
Metrics
Altmetric Badge
Dimensions Badge
Share
Actions (login required)
View Item |