AL-SAGUR, Hadi Wali Saheb (2018). Development of glucose biosensors using nanostructured composites. Doctoral, Sheffield Hallam University. [Thesis]
Documents
23405:520828
PDF
Al-Sagur_2018_PhD_DevelopmentOfGlucose.pdf - Accepted Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.
Al-Sagur_2018_PhD_DevelopmentOfGlucose.pdf - Accepted Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.
Download (8MB) | Preview
Abstract
Developing a biosensor capable of measuring glucose and other whole blood analytes for monitoring diabetes has been a major challenge for over four decades. In this thesis, an attempt has been made to develop three different novel metallophthalocyanine-based (MPcs) biosensors for this purpose. Three novel enzymatic biosensors have been fabricated to monitor glucose concentration and other whole blood analytes in vitro. The first fabricated biosensor is based on conducting multifunctional hydrogel (PAA-rGO/VS-PANI/LuPc2/GOx-MFH) utilising reduced graphene oxide (rGO) and lutetium phthalocyanine (LuPc2) dissolved in chloroform to produce three-dimensional (3D) matrix as platform to enhance the sensing performance. Utilising the aqueous properties of a novel water-soluble iron phthalocyanine (FePc) derivative on the other hand, added more simplicity for the fabrication process of another biosensor (PAA-CP/GPL-FePc/GOx-CH) for enzymatic detection of glucose. Graphene nanoplatelets (GPL) have been attached non-covalently to the FePc resulting in a conducting hydrogel-based platform (CH). A third novel bioprobe (SiO2(LuPc2)-PANI(PVIA)/GOx-CNB) based on silica nanoparticles (SiO2) grafted polyaniline (PANI) has formed a conducting nanobeads (CNB)-based biosensor. The latter is employed as an enzymatic biosensor platform for the detection of glucose with enhanced sensitivity. Full characterisation has been carried out for all the raw studied materials as well as prepared biosensing platforms. UV-Visible and FT-IR spectroscopies as well as SEM, TEM, XRD, and EDX have been employed in order to help gaining full understanding of the nature and properties of the studied materials. The electrochemical properties of the newly developed biosensing platforms have been fully studied using common analytical methods such as cyclic voltammetry (CV), amperometry, and electrochemical impedance spectroscopy (EIS). The freeze dry system alongside the Brunauer-Emmett-Teller (BET) method were employed to characterise the surface area of the produced platforms.
The sensitivity of MFH biosensor studied in the range 2-12 mM of glucose is found to fall in the region of 15.31 μA mM−1 cm−2 with low detection limit of 25 μM. The biosensor based on CH platform exhibited a broad linear behaviour when glucose in the range 1-20 mM is studied, with high sensitivity of 18.11 μA mM−1 cm−2 and low detection limit of 1.1503 ng/mL. The conducting nanobeads-based biosensor (CNB) has led to a further improvement the sensitivity of glucose detection (38.53 μA mM−1 cm−2) with wide linear range of 1-16 mM and detection limit of 0.1 mM. All three biosensing platforms have exhibited excellent selectivity when examined against whole blood components.
More Information
Statistics
Downloads
Downloads per month over past year
Metrics
Altmetric Badge
Dimensions Badge
Share
Actions (login required)
View Item |