Effect of limb surrogate surface compliance on the impact response of wrist protectors

LESLIE, Gemma E, WINWOOD, Keith, WANG, Weizhuo, HAMILTON, Nick and ALLEN, Tom (2023). Effect of limb surrogate surface compliance on the impact response of wrist protectors. JSAMS Plus, 2: 100023.

[img]
Preview
PDF
Leslie-EffectOfLimbSurrogateSurface(VoR).pdf - Published Version
Creative Commons Attribution.

Download (1MB) | Preview
Official URL: https://www.sciencedirect.com/science/article/pii/...
Open Access URL: https://doi.org/10.1016/j.jsampl.2023.100023 (Published version)
Link to published version:: https://doi.org/10.1016/j.jsampl.2023.100023

Abstract

Objectives To investigate the effect on the impact response of wrist protectors by incorporating a soft tissue simulant on to a wrist surrogate made of stiff material. The effect of protector strapping condition was also investigated. Design and methods A compliant and a stiff surrogate were made, based on the wrist geometry specified in ISO 20320:2020 for the “Limitation of wrist extension” test. Two styles of wrist protectors (short, long) were tested on each limb surrogate, subject to a ∼31 ​J impact. Six protectors of each style were tested, with two of each at each strapping condition (loose, moderate, tight) on each surrogate (stiff, compliant) (24 combinations). Example temporal force and wrist angle and force vs. wrist angle plots are presented for comparison between conditions. Results When protectors were on the compliant surrogate, peak impact force was 55–68% lower (short 3.1 vs. 6.8 ​kN, long 2.7 vs. 8.3 ​kN). The time to reach this peak force was ∼4 ​ms (12%) longer, than for the stiff surrogate. Protector strapping condition had no clear effect for the stiff surrogate, with the wrist extending to its limit for all tests. Strapping protectors tighter on the compliant surrogate tended to decrease the maximum wrist angle and peak force. Conclusions With results being sensitive to surrogate design and strapping condition, these both need to be clearly reported in future work impact testing wrist protectors, with implications for certification tests within standards.

Item Type: Article
Identification Number: https://doi.org/10.1016/j.jsampl.2023.100023
SWORD Depositor: Symplectic Elements
Depositing User: Symplectic Elements
Date Deposited: 16 May 2023 10:03
Last Modified: 11 Oct 2023 15:00
URI: https://shura.shu.ac.uk/id/eprint/31872

Actions (login required)

View Item View Item

Downloads

Downloads per month over past year

View more statistics