B-Mode Ultrasonography Is a Reliable and Valid Alternative to Magnetic Resonance Imaging for Measuring Patellar Tendon Cross-Sectional Area

MARSHALL, Steven J, HICKS, Kirsty M, HOWATSON, Glyn, COX, Lisa and THOMAS, Kevin (2022). B-Mode Ultrasonography Is a Reliable and Valid Alternative to Magnetic Resonance Imaging for Measuring Patellar Tendon Cross-Sectional Area. Ultrasound in Medicine & Biology, 49 (2), 578-587.

[img]
Preview
PDF
Marshall-B-ModeUltrasonographyIsAReliableAndValid(VoR).pdf - Published Version
Creative Commons Attribution.

Download (1MB) | Preview
Official URL: https://www.sciencedirect.com/science/article/pii/...
Link to published version:: https://doi.org/10.1016/j.ultrasmedbio.2022.10.011

Abstract

This study investigated the validity and reliability of measuring patellar tendon (PT) cross-sectional area (CSA) using magnetic resonance imaging (MRI) and ultrasound (US) imaging. Nineteen healthy participants (10 women, 9 men) participated in three imaging sessions of the PT, once via MRI and twice via US, with image acquisition conducted by two raters, one experienced (rater 2) and one inexperienced (rater 1). All PT segmentations were analyzed by both raters. The validity of US-derived estimates of PT CSA against MRI estimates was analyzed using linear regression. Within-day reliability of US and MRI measurements and between-day reliability of US measurements were quantified using typical error (TE) and intra-class correlation coefficients (ICC3,1). There was good agreement between US- and MRI-derived estimations of PT CSA (standard errors of the estimate of 3.3 mm2 for rater 1 and 2.6 mm2 for rater 2; Pearson's r = 0.97 and 0.98 for raters 1 and 2, respectively). Within-session reliability for estimations of total PT CSA from US and MRI were excellent (ICC3,1 >0.95, coefficient of variation [CV] <4.1%, TE = 1.3–3.6 mm2. Between-day reliability for US was excellent (ICC3,1 >0.97, CV <2.7%, TE = 1.6–2.3 mm2), with little difference between raters. These findings suggest that MRI and US both provide reliable estimates of PT CSA and that US can provide a valid measure of PT CSA.

Item Type: Article
Uncontrolled Keywords: Acoustics; 1103 Clinical Sciences
Identification Number: https://doi.org/10.1016/j.ultrasmedbio.2022.10.011
Page Range: 578-587
SWORD Depositor: Symplectic Elements
Depositing User: Symplectic Elements
Date Deposited: 15 Dec 2022 08:35
Last Modified: 12 Oct 2023 08:16
URI: https://shura.shu.ac.uk/id/eprint/31176

Actions (login required)

View Item View Item

Downloads

Downloads per month over past year

View more statistics